Pulse amplitude adjustment provides immediate pacemaker longevity gain

Journal
Anadolu kardiyoloji dergisi : AKD = the Anatolian journal of cardiology
Date Issued
2007-07
Author(s)
Zlatanovic, Nenad
Gjorgov, Nikola
Miletic, Branislav
Kovacevic, Dejan
Kjaev, Mitko
Borozanov, Vladimir
Boshkov, Vladimir
Abstract
Objective: Adjusting pacemaker pulse amplitude influences the longevity of the pacemaker. Our aim was to establish the initial longevity gain. Methods: Forty randomly selected patients with implanted pacemakers were analyzed. Mean age was 65.58±13.7 years. All pacemakers were working on factory settings of pulse amplitude 3.5V and pulse width of 0.4 ms for average of 3 years before the adjustment. Initial mean longevity was projected to 68.61±18.86 months, mean battery voltage 2.78V, and mean battery current 14.21±2.61 µA. Results: Pulse amplitude threshold test was performed and average value of 0.632±0.22V was obtained. Pulse amplitude was programmed to 2.5V and pulse width was left unchanged. New readings of battery data were obtained. Battery voltage did not show immediate changes, and battery current decreased to 11.53±1.98 µA . New average longevity was projected to 81.03±19.82 months, which presents a 12.42 months of initial longevity gain with statistical significance at 95% confidence interval (p=0.003). Positive correlation was found between the new pulse amplitude and new values of battery current (p<0.01). Conclusion: Pulse amplitude decrease of only 1V provides significant initial longevity gain of more than a year. If found correlations would have any impact on further longevity gains over longer period of time is yet to be established.
Subjects

pacemakers

longevity

pulse amplitude

pulse width

pacing threshold

File(s)
Loading...
Thumbnail Image
Name

AnatolJCardiol_7_1_216_218.pdf

Size

48.66 KB

Format

Adobe PDF

Checksum

(MD5):b97effd0b47e3c4f1468ef24de4bea59

⠀

Built with DSpace-CRIS software - Extension maintained and optimized by 4Science

  • Accessibility settings
  • Privacy policy
  • End User Agreement
  • Send Feedback
Repository logo COAR Notify